Fabrication of TlBr strip detectors
Hitomi, K, Kikuchi, Y, Nakhostin, M, Shoji, T and Ishii, K (2009) Fabrication of TlBr strip detectors In: Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XI, 2009-08-03 - 2009-08-06, San Diego, USA.
Full text not available from this repository.Abstract
TlBr strip detectors with four anode strip electrodes were fabricated in this study. The strip electrodes were formed on TlBr wafers 1 mm thick with a vacuum evaporation process through a shadow mask. Each strip had a width of 0.8 mm and a length of 3.5 mm approximately. The gap between the strips was approximately 0.1 mm. Inter strip resistance more than 1 GΩ was obtained with the process. Energy resolutions of 7.0% and 3.8% FWHM were recorded for 511 keV and 1.33 MeV gamma-rays, respectively, with the TlBr strip detector at room temperature.