University of Surrey

Test tubes in the lab Research in the ATI Dance Research

Quantifying nitrogen in GeSbTe:N alloys

Jeynes, C., Nolot, E., Costa, C., Sabbione, C., Pessoa, W., Pierre, F., Roule, A. and Mantler, M. (2018) Quantifying nitrogen in GeSbTe:N alloys Journal of Analytical Atomic Spectrometry.

[img] Text
GSTN-JAASsubmissionR1.docx - Accepted version Manuscript
Restricted to Repository staff only

Download (1MB)

Abstract

We have calibrated on-site WD-XRF (wavelength-dispersive X-ray fluorescence) measurements of GeSbTe:N (GST:N) stoichiometry with off-site accurate ion beam analysis (IBA). N is determined by elastic backscattering spectrometry (EBS) using the resonance at 3.7 MeV in the 14N(a, a)14N reaction. Ge and Sb+Te are determined by Rutherford backscattering spectrometry (RBS) separately but self-consistently with the resonant EBS: the Sb/Te ratio can be determined by RBS but not with useful precision. The XRF instrumental function is determined using pure metal standards and the spectra are quantified using Fundamental Parameters code. We find that, as expected, for both Ge and (Sb+Te) the heavy elements are determined accurately by XRF (within the uncertainties), but for N the standardless XRF has non-linear errors around 10%. Using the absolute N content determined by IBA a calibration curve is obtained allowing N determination by WD-XRF at a precision of about 1% and an absolute accuracy (traceable through IBA) of about 4 % for GST films with N content between 4-20 at%. The IBA measurement precision of the N content of the GST-N XRF calibration samples is 0.4 at% (that is, a relative precision ranging from 10 % to 2 % for N contents between 4-20 at%).

Item Type: Article
Divisions : Faculty of Engineering and Physical Sciences > Electronic Engineering
Authors :
NameEmailORCID
Jeynes, C.c.jeynes@surrey.ac.uk
Nolot, E.
Costa, C.c.d.costa@surrey.ac.uk
Sabbione, C.
Pessoa, W.
Pierre, F.
Roule, A.
Mantler, M.
Date : 3 December 2018
Copyright Disclaimer : © The Royal Society of Chemistry 2020
Uncontrolled Keywords : Accurate IBA; Certified reference materials; Fundamental Parameters method; Industrial metrology; Ion beam analysis; Resonant elastic scattering; Phase change materials; Quality control; PCRAM
Depositing User : Diane Maxfield
Date Deposited : 10 Jan 2020 12:54
Last Modified : 10 Jan 2020 12:54
URI: http://epubs.surrey.ac.uk/id/eprint/853308

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year


Information about this web site

© The University of Surrey, Guildford, Surrey, GU2 7XH, United Kingdom.
+44 (0)1483 300800