Finite element analysis: a comparison of an all-polyethylene tibial implant and its metal-backed equivalent.

Thompson, SM, Yohuno, D, Bradley, WN and Crocombe, AD (2015) Finite element analysis: a comparison of an all-polyethylene tibial implant and its metal-backed equivalent. Knee Surg Sports Traumatol Arthrosc. pp. 1-7.

[img] Text
KSST for SRI.pdf - Accepted version Manuscript
Restricted to Repository staff only until 22 December 2016.
Available under License : See the attached licence file.

Download (806kB)
[img]
Preview
Text (licence)
SRI_deposit_agreement.pdf
Available under License : See the attached licence file.

Download (33kB) | Preview

Abstract

PURPOSE: The hypothesis of this study is that all-polyethylene (APE) tibial implants offer a biomechanical profile similar to metal-backed tray (MBT). There are significant financial implications, in selected patient groups, if APE can be deemed to perform as well as MBT. METHODS: Using a finite element analysis of CAD models provided by DePuy (Leeds), stress distributions were investigated for both an APE and MBT tibial implant. The performance was assessed for cancellous bone at 700 MPa (normal) and at 350 MPa (less stiff). Plots were recorded along the length of the tibia, showing the loads carried by the bone (cortical and cancellous), the implant interface, cement interface and the stem. von Mises stress distributions and percentage volumes were used to assess bone resorption and hence potential for failure (fracture). RESULTS: Higher stress shielding (resorption) occurred around the keel and stem of the MBT revealing greater potential for bone loss in these areas. APE had no areas of bone resorption (being more flexible resulting in less stress shielding). The stiffer MBT carries a higher proportion of the load down the stem. MBT stress in cancellous bone is lower than APE, as load is distributed to the cortical rim. APE has a marginally favourable strain state in cancellous bone and spreads loads more at the cement interface than MBT. CONCLUSION: Modern-day APE bearings may be superior to previously designed implants due to improvements in manufacturing. In the correct patient group, this could offer substantial cost savings.

Item Type: Article
Subjects : Mechanical Engineering Sciences
Divisions : Faculty of Engineering and Physical Sciences > Mechanical Engineering Sciences
Authors :
AuthorsEmailORCID
Thompson, SMUNSPECIFIEDUNSPECIFIED
Yohuno, DUNSPECIFIEDUNSPECIFIED
Bradley, WNUNSPECIFIEDUNSPECIFIED
Crocombe, ADUNSPECIFIEDUNSPECIFIED
Date : 22 December 2015
Identification Number : 10.1007/s00167-015-3923-y
Copyright Disclaimer : The final publication is available at Springer via http://dx.doi.org/[10.1007/s00167-015-3923-y
Uncontrolled Keywords : All-polyethylene tibial implant, Finite element analysis, Metal-backed tibial tray, Stress
Related URLs :
Additional Information : The final publication is available at Springer via http://dx.doi.org/[10.1007/s00167-015-3923-y
Depositing User : Symplectic Elements
Date Deposited : 02 Mar 2016 10:22
Last Modified : 02 Mar 2016 10:22
URI: http://epubs.surrey.ac.uk/id/eprint/809994

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year


Information about this web site

© The University of Surrey, Guildford, Surrey, GU2 7XH, United Kingdom.
+44 (0)1483 300800